When using transcutaneous CO2 monitoring, whether in the NICU, PICU, sleep lab, or even at home, sensor performance is critical to delivering reliable data. A well-functioning sensor helps ensure the accuracy clinicians depend on to make informed care decisions.

Regular maintenance plays a key role in keeping your sensors performing at their best, supporting consistent, accurate monitoring and confidence in every reading.

FOR YOUR TEAM

Proactive Maintenance Checklist

Make sensor maintenance simpler with a step-by-step schedule designed to support team consistency and optimal sensor performance.

Cleaning Your Sensor

It is recommended to clean your sensor after each use to remove residue and buildup, and perform a deeper, quarterly clean (Clean and Soak) to maintain optimal performance. Follow the steps below to keep your sensor clean:

After-Use Cleaning

In the short term, painful procedures can trigger measurable physiological stress responses that may affect clinical stability. During painful events, infants may experience:

  1. Remove the sensor from the patient and inspect the skin.
  2. Clean the sensor with a gauze soaked in 70% isopropanol or other approved cleaning agent.
  3. Gently wipe the sensor surface, then clean the sides of the membrane ring.
  4. Avoiding strain on the cable, gently clean the sensor cable from the sensor head to plug.
  5. Visually inspect the sensor and membrane.
  6. Place the sensor in the Docking Station and allow it to calibrate.
  7. Once you see the “Ready for Use” screen on the monitor, you can start a new monitoring session.

Clean and Soak

  1. Disconnect the sensor from the sensor adapter cable.
  2. Remove the current membrane from the sensor using the bottom of the membrane changer. If the membrane does not come off easily, you can first soak the sensor in clean water with the membrane on.
  3. Once the membrane is removed, immerse the sensor in clean room temperature water for up to 3 minutes.
  4. After soaking, gently rinse the sensor with clean water.
  5. Dry the grooves around the sensor with clean, lint-free gauze, making sure there is no gel residue remaining. To dry the sensor surface, gently tap with gauze. (Do not wipe or rub the sensor surface, as this can cause damage.)

Inspecting Your Sensor

Before and after each use, inspect your sensor to ensure it is in good working condition. Check for the following:

  • The membrane is present and intact
  • No air bubbles are visible beneath the sensor membrane
  • The area under the membrane is dark gray, indicating the electrolyte solution is in good condition (a lighter color or cloudy appearance may suggest the solution has dried out)
  • The silver chloride ring (the innermost circle on the sensor) should appear brown. If the center ring has a white or silver luster, the sensor may be defective or deteriorated

  • The LED light glows red when connected to the monitor
  • There is no visible wear on the sensor head or cable

Changing the Sensor Membrane

The Sentec sensor contains a thin layer of electrolyte on its surface, covered by a CO2-permeable membrane. Over time, the electrolyte gradually dries out, making periodic membrane replacement necessary.

A membrane change should be performed on the sensor monthly using a Membrane Changer, following these simple steps:

  1. Place the sensor in the Membrane Changer.
  2. Complete the four pressing and turning steps.
  3. Remove the sensor and inspect the membrane.
  4. Replace the sensor in the Docking Station and confirm the membrane change on the monitor.

Performing a Sensitivity Test

Another important part of your monthly maintenance routine is performing a Sensitivity Test, which helps ensure your sensor is monitoring accurately. To perform a Sensitivity Test, follow these steps:

  1. Navigate to the Sensitivity Test
    1. On the SDM, enter the Main Menu, then select ‘Measurement Settings’, ‘PCO2 Settings’, and then ‘Sensitivity Test’.
    2. On the tCOM+, enter the Main Menu, then select ‘Sensor Maintenance’ and then ‘Sensitivity Test’.
  2. Open the door to expose the sensor to ambient air, as prompted by the monitor.
  3. After 2 minutes, the monitor will prompt you to close the door.
  4. Once the door is closed, the monitor will run an automatic Calibration and Leak Test.
  5. If the sensor passes the Sensitivity Test, the message ‘Ready for use’ will appear.

Troubleshooting Error Messages

A well-planned maintenance routine helps support consistent performance and reliability of your transcutaneous sensors. However, occasional error messages can still occur.

Explore the resources below to learn how to troubleshoot sensor errors, so you can quickly resolve issues and minimize interruptions in monitoring.

How to resolve sensor problem 11

How to resolve sensor problem 12

Maintenance & Troubleshooting Tip Sheet

Need support?

Our Product Support team is ready to support you at every step of using Sentec transcutaneous CO2 monitoring. Reach out to us anytime or browse our product support page for tools and resources tailored to your team.

Related Posts

  • Whitepaper

    Transcutaneous CO2 Monitoring in the NICU: Practical Applications, Protocols, and Tips for Successful Implementation

    Anne Geistkemper, MSc, RRT, RRT-NPS, shares how Rush University Children's Hospital integrates transcutaneous CO₂ monitoring into NICU care to support ventilation management, reduce blood draws, minimize neonatal pain, and improve clinical decision-making.

  • Whitepaper

    Transcutaneous CO2 Monitoring for Earlier Interventions in the PICU: Case Studies from Children’s Memorial Hermann

    Learn how Children's Memorial Hermann uses transcutaneous CO₂ monitoring in the PICU to support early intervention and ventilatory management through four real-world pediatric case studies.

  • Whitepaper

    Balancing Brain & Lung Protection in the NICU with Continuous CO2 Monitoring

    Continuous CO2 monitoring helps NICU teams balance neuroprotective and lung-protective care in premature infants by providing real-time visibility to ventilation status, reducing reliance on blood draws, and supporting individualized clinical decision-making.